Atomic absorption spectroscopy
Atomic absorption spectroscopy (AAS) is a spectroanalytical procedure for the quantitative determination of chemical elements using the absorption of optical radiation (light) by free atoms in the gaseous state.
Atomic absorption spectrometry has many uses in different areas of chemistry such as:
- Clinical analysis: Analyzing metals in biological fluids and tissues such as whole blood, plasma, urine, saliva, brain tissue, liver, muscle tissue, semen
- Pharmaceuticals: In some pharmaceutical manufacturing processes, minute quantities of a catalyst that remain in the final drug product
- Water analysis: Analyzing water for its metal content.
Atomic Absorption Theory
Atomic absorption spectroscopy relies on the Beer-Lambert law to determine the concentration of a particular analyte in a sample. The absorption spectrum and molar absorbance of the desired sample element are known, a known amount of energy is passed through the atomized sample, and by then measuring the quantity of light it is possible to determine the concentration of the element being measured.
Instrumentation
In order to analyze a sample for its atomic constituents, it has to be atomized. The atomizers most commonly used nowadays are flames and electrothermal (graphite tube) atomizers. The atoms should then be irradiated by optical radiation, and the radiation source could be an element-specific line radiation source or a continuum radiation source. The radiation then passes through a monochromator in order to separate the element-specific radiation from any other radiation emitted by the radiation source, which is finally measured by a detector.
Atomizers
The atomizers most commonly used nowadays are (spectroscopic) flames and electrothermal (graphite tube) atomizers. Other atomizers, such as glow-discharge atomization, hydride atomization, or cold-vapor atomization might be used for special purposes.
Flame atomizers
The oldest and most commonly used atomizers in AAS are flames, principally the air-acetylene flame with a temperature of about 2300 °C and the nitrous oxide system (N2O)-acetylene flame with a temperature of about 2700 °C. The latter flame, in addition, offers a more reducing environment, being ideally suited for analytes with high affinity to oxygen.
Electrothermal atomizers
Although a wide variety of graphite tube designs have been used over the years, the dimensions nowadays are typically 20–25 mm in length and 5–6 mm inner diameter. With this technique liquid/dissolved, solid and gaseous samples may be analyzed directly. A measured volume (typically 10–50 μL) or a weighed mass (typically around 1 mg) of a solid sample are introduced into the graphite tube and subject to a temperature program. This typically consists of stages, such as:
- Drying – the solvent is evaporated
- Pyrolysis – the majority of the matrix constituents are removed
- Atomization – the analyte element is released to the gaseous phase
- Cleaning – eventual residues in the graphite tube are removed at high temperature.